Minimising errors in low resistance measurements

Power Parameters Pty Ltd
Wednesday, 05 January, 2005


This article looks at variables in measuring low resistance values and the errors that may occur in making those measurements.

Two-wire versus 4-wire connection to the device under test (DUT) is an important factor in obtaining accurate results.

Another consideration is the thermal EMF produced at the junctions of dissimilar metals that can cause decreased accuracy at very small voltage levels. Device heating and inductive devices may add to the EMF error.

Dry circuit testing may be implemented to avoid oxidation puncture of the test leads. Determining temperature rise in the self-heating of motors, transformers, solenoids and coils provides motor and transformer manufacturers with a more accurate characterisation of their device.

Most of today's digital multimeters (DMM) and some dedicated resistance measurement instruments use a 2-wire test method. In this, the test current (ITEST) is forced through the test leads and across the resistance (RDUT) being measured.

The meter then measures the voltage across the resistance through the same set of leads and the resistance value is computed.

Figure 1: 2-wire connection to DUT.

The resistance of the test lead (RLEAD) is the concern with the 2-wire method when making low resistance measurements. The test current (ITEST) causes a small yet significant voltage drop across the lead resistances.

The voltage drop (VMETER) measured by the meter will not be exactly the same as the voltage (VDUT) directly across the device under test (RDUT) and considerable errors can result. Typical lead resistances commonly range from 0.01-1 ohm making accurate 2-wire measurements below 10 ohm difficult to obtain.

Due to the limitations of the 2-wire method, the 4-wire (Kelvin) connection is implemented in most milliohmmeters. With a 4-wire configuration, one set of leads drives the current (or voltage) and the second set of leads senses the voltage (or current).

Figure 2 illustrates the 4-wire connection to DUT. Here, the test current (ITEST) is forced through the DUT (RDUT) through one set of leads, while the voltage across the DUT (VDUT) is measured by a second set of leads.

Figure 2: 4-wire connection to DUT.

Although some small current may flow through the voltage leads it is usually small enough to be ignored. Since the voltage drop across the voltmeter leads is negligible, the voltage across the meter can be considered the voltage across the DUT.

In essence the resistance of the DUT (RDUT) can be measured more accurately with the 4-wire method. The voltage sensing leads should be connected as close as possible to the DUT to avoid including the effects of the voltage drop across the test leads in the final measurement.

Thermal EMFs are small voltages developed at the junctions of dissimilar metals. The magnitude of thermal EMF depends on both the type of metal used and the temperature difference between the junctions.

Since low resistance measurements are dependent on the test instrument's ability to measure very low voltage levels, thermal EMF can significantly contribute to low resistance measurement error.

External to the DUT, each connection or connector in a test setup is a possible thermal EMF source. These include connections between the DUT and the input cables; connections within the input cables and connections between the input cable and the instrument's input connector.

Even connections internal to the instrument can cause thermal EMF. Thermal EMF sources external to the DUT can be cancelled out by the zeroing function of most of today's instrumentation, like the QuadTech LR2000 milliohmmeter.

Simple instrument zeroing will not compensate for the thermal EMF sources associated with connections within the DUT or in other connections beyond the instrument's input terminals (the point at which zeroing is performed).

Figure 3: Current reversal method.

Suggested techniques for minimising thermal EMF include using only clean crimped-on similar metal (copper to copper) connections and keeping all junctions at the same temperature. This is not practical in all test applications so there are two common methods that are used in many milliohmmeters to circumvent this problem. The two methods are current reversal and offset-compensated ohms.

Using the current reversal method, thermal EMF is cancelled by making two measurements with currents of opposite polarity. The positive current (+ISOURCE) is applied and the voltage is measured (VMETER+).

A negative current (-ISOURCE) is applied and the voltage is measured a second time (VMETER-). The two measurements are then combined to cancel any effect of thermal EMF. Refer to Figure 3 for equations.

The measured resistance is then computed using Ohm's Law as RMEAS = VSOURCE/ISOURCE. Figure 3 illustrates the current reversal method.

The offset-compensated ohm method for minimising thermal EMF applies the source current (ISOURCE) to the resistance being measured (RMEAS) only during one part of the test cycle.

When the source is ON, the total voltage measured (VMETER 1) includes the resistor as well as any thermal EMF as illustrated in Figure 4.

The second voltage measurement (VMETER 2) is made with the current source OFF. The two voltage measurements are then combined to determine the voltage measurement for the full test cycle. This voltage is termed the offset-compensated voltage.

Device heating may be a source for error in making low resistance measurements on temperature sensitive devices like thermistors. Due to the high test current being supplied by the milliohmmeter, the power dissipated by the DUT can cause heat spikes large enough to alter the DUT's resistance.

Although the test current is not adjustable in most milliohmmeters there is a way of addressing the problem of DUT heating. A triggered measurement mode in which a single momentary current pulse is applied to the DUT during the measurement cycle to reduce errors caused device heating.

Figure 4: Offset-compensated ohms method.

Although pulsing the test current provides the benefits of compensating for thermal EMF and minimising device heating, current pulsing may cause errors in testing inductive devices. The inductance of the DUT may prevent the current through the device from reaching its maximum value before the voltage measurement is made.

This phenomenon is due to the L/R time constant being larger than the current pulse width.

When this is the case the current never reaches its maximum value and the measured voltage is too low, which results in an erroneous resistance measurement. The solution is to use 'straight', non-pulsed DC test current when testing an inductive device. Take into consideration the previous discussion that such test currents could produce device heating depending on the DUT.

The QuadTech LR2000 millohmmeter is a solution for most inductive device measurement applications.

Many low resistance measurements are made on contacts of devices such as switches and relays. This test can determine if oxidation has increased the resistance of these contacts. If the voltage across the contacts is too high the oxidation layer will be punctured and the validity of the test is then compromised.

Figure 5: Inductive DUT.

To avoid oxidation puncture these measurements are usually made using 'dry circuit' testing. The open circuit voltage across the instrument's test leads is held constant (clamped) at 20 mV or less. The LR2000 milliohmmeter features dry circuit mode for this purpose.

The determination of the temperature rise in motors and transformers due to self-heating is a very common measurement. Motors, transformers, solenoids and coils all exhibit symptoms of heat rise during use.

The internal power losses of the device result in heating which increases the operating temperature of the unit. In most cases it is impractical to measure the temperature with thermocouples or other temperature sensors, hence the change in resistance method for temperature determination.

The majority of magnetic devices use either copper or aluminium wire in the construction of their core. These wires have precise temperature coefficients (TC) which can be used with resistance measurements to calculate the temperature rise (DT) of the device under test (DUT).

The change in temperature is equal to the resistance of the DUT before use (RCOLD) minus the resistance of the DUT during use (RHOT) divided by the temperature coefficient times RCOLD.

Let's look at the calculation of temperature rise for a motor after eight hours of operation at a specified rated load. The field winding of the motor is copper wire that has a temperature coefficient of 3931 ppm/°C (DR = 0.3931%/°C).

Before running at a load the ambient winding resistance is measured as 1.2367 ohm. After eight hours at full load, the winding resistance is measured as 1.6211 ohm. Therefore, the computation of the temperature rise is:

Ambient temperature changes could have significant impacts on the test results. Some milliohmmeters have a temperature sensing function to measure the ambient temperature or capability for entering this data. The test results and temperature conditions are then automatically referenced to nominal ambient temperature (23°C).

QuadTech has created a Visual Basic application wizard for the LR 2000 that calculates temperature compensation. Just enter the temperature and select your material (or enter your coefficient) and both resistance values are displayed and logged with a click on the [trigger] button.

The LR2000 milliohmmeter - a compact, portable unit is designed for accurate low resistance measurements on switches, relays, connectors and cables. The unit provides eight automatic or manually selectable measurement ranges from 20 mohm to 2 Mohm with constant current ranging from 1 A to 1 µA.

For remote operation and production applications, the milliohmmeter has an RS232 interface as standard and an optional IEEE 488/handler interface.

The instrument has comparator, binning and remote functions making it suitable for Go/NoGo testing and it provides four-terminal Kelvin connection to the DUT making accurate low resistance measurements down to 1 µohm possible.

Related Articles

Hidden semiconductor activity spotted by researchers

Researchers have discovered that the material that a semiconductor chip device is built on,...

3D reflectors help boost data rate in wireless communications

Cornell researchers have developed a semiconductor chip that will enable smaller devices to...

Scientists revolutionise wireless communication with 3D processors

Scientists have developed a method for using semiconductor technology to manufacture processors...


  • All content Copyright © 2024 Westwick-Farrow Pty Ltd